Surgical Nursing Free guide

Sterilization Methods, Indicators and Pack Storage for Veterinary Technicians

Surgical Nursing (D2 I; K18, K19) 13 min read, 2,781 words Last verified

How to choose and run a sterilization method, read chemical and biological indicators, load the autoclave and store packs under event-related shelf life.

Introduction

Sterilization is the process that destroys or removes all forms of microbial life, including bacterial spores.1 In most veterinary hospitals the technician owns this job. You clean the instruments, wrap the packs, load and run the autoclave, read the indicators, keep the logs and decide whether a pack is safe to open.

This guide covers the Spaulding categories, steam sterilization (gravity displacement and prevacuum), ethylene oxide, hydrogen peroxide gas plasma, liquid chemical agents, the indicator classes, biological indicators, autoclave loading, event-related shelf life and basic autoclave checks. Each section focuses on what you do and what you check.

Key facts

  • Cleaning comes first. Organic and inorganic soil left on an item shields microbes and makes disinfection or sterilization fail.1
  • Critical items (anything that enters sterile tissue or the vascular system) must be sterile. Semicritical items touch mucous membranes or nonintact skin. Noncritical items touch only intact skin.1
  • Recognized minimum steam exposure for wrapped supplies: 30 minutes at 121°C (250°F) in a gravity displacement sterilizer, or 4 minutes at 132°C (270°F) in a prevacuum sterilizer.1
  • Ethylene oxide (EO or ETO) must be followed by aeration: 8 to 12 hours of mechanical aeration at 50 to 60°C, or 7 days at 20°C room air.1
  • Hydrogen peroxide gas plasma cannot process cellulose (paper), linens or liquids.1
  • Biological indicators: Geobacillus stearothermophilus spores for steam, gas plasma and peracetic acid; Bacillus atrophaeus spores for ethylene oxide and dry heat. Run them at least weekly.1
  • Indicator tape on the outside of a pack shows only that the pack went through a cycle. It does not prove sterility.13
  • Under event-related shelf life, a pack stays sterile until something happens to it: it gets wet, torn, punctured or the seal breaks.1

How the VTNE tests this

  • Choose the method for an item: steam for heat-stable metal, low-temperature gas for heat- or moisture-sensitive items (Domain 2 Task I; K18).6
  • Interpret an indicator: what a color-changed tape, an internal strip, an integrator or a spore test result means (K18).
  • Recall cycle numbers: 121°C for 30 minutes (gravity) versus 132°C for 4 minutes (prevacuum).
  • Judge a stored pack: is it wet, torn or punctured, and what do you do next.
  • Recognize hazards: ethylene oxide exposure, aldehyde irritation, burns from hot instruments (K19).6

Core concepts

Cleaning, disinfection and sterilization

Cleaning is the removal of visible soil, such as blood, tissue and debris, usually with water and a detergent or enzymatic product.1 It does not kill everything, but it lowers the number of organisms the next step must kill. In one classic test, killing 10 B. atrophaeus spores took 30 minutes, while killing 100,000 spores under the same conditions took 3 hours.1

Disinfection eliminates many or all pathogenic microorganisms on objects, except bacterial spores.1 High-level disinfection kills all microorganisms except large numbers of bacterial spores. A few liquid disinfectants kill spores if the item stays immersed long enough (3 to 12 hours). Used that way they are called chemical sterilants. Used for a short time, such as 20 minutes for 2% glutaraldehyde, the same product only achieves high-level disinfection.1

Sterilization destroys all microbial life. Steam under pressure, dry heat, ethylene oxide, hydrogen peroxide gas plasma and liquid chemicals are the main sterilizing agents in health care.1

Spaulding categories

The Spaulding system sorts items by infection risk, and the risk sets the level of processing.

Table 1. Spaulding categories and the processing each needs
CategoryContactVeterinary examplesMinimum processing
CriticalSterile tissue or the vascular systemSurgical instruments, implants, urinary catheters, extraction forceps, surgical bursSterilization
SemicriticalMucous membranes or nonintact skinLaryngoscope blades, anesthesia and respiratory equipment, some endoscopesHigh-level disinfection at least; sterilize if heat tolerant
NoncriticalIntact skin onlyBlood pressure cuffs, exam tables, computer keyboardsCleaning and low-level disinfection

Source: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities.1

Steam sterilization

Moist heat kills by irreversible coagulation and denaturation of enzymes and structural proteins.1 Steam sterilization has four parameters: steam, pressure, temperature and time. Pressure is not what kills. It raises the boiling point so the steam reaches 121°C or 132°C.1 Steam is the preferred method for any instrument not damaged by heat, steam, pressure or moisture.1 It is also the most common method in veterinary practice.2 Oils and greases cannot be steam sterilized.2

The enemy of steam is trapped air. Air pockets keep steam from touching the surfaces inside a pack.

  • Gravity displacement autoclaves let steam in at the top and push the cooler, heavier air out through a drain at the bottom. Air leaves slowly from porous loads, so cycles are long.1 These are the most common type in private practice.2
  • Prevacuum (high-speed prevacuum) autoclaves use a vacuum pump to remove air from the chamber and load before steam enters. Steam penetration, even into porous packs, is nearly instant.1
  • Steam-flush pressure-pulse units remove air by alternating steam flushes and pressure pulses.1
Air removal in gravity displacement and prevacuum steam autoclaves Gravity displacement121°C (250°F), 30 min wrappedSteam enters at topAir pushed out the drainHeavier, cooler air sinksPrevacuum132°C (270°F), 4 min wrapped1. Pump pulls air outtopump2. Steam fills all spacesNear-instant pack penetrationDashed boxes are wrapped packs. Trapped air blocks steam contact.
Figure 2. Air removal decides cycle length. A gravity displacement autoclave pushes air down and out the drain, so wrapped loads need 30 minutes at 121°C (250°F). A prevacuum autoclave pumps the air out first, so steam reaches the pack center at once and 4 minutes at 132°C (270°F) is enough.

Low-temperature methods

Ethylene oxide is a colorless gas that is flammable and explosive. It kills by alkylation of protein, DNA and RNA.1 Its operating ranges are a temperature of 37 to 63°C, relative humidity of 40 to 80% and exposure of 1 to 6 hours. A basic cycle takes about 2.5 hours before aeration starts.1 Absorbent materials hold toxic EO residue, so aeration is not optional. Contact with items that still hold EO can burn patients and handlers. Plastics and rubber need much longer aeration than stainless steel.2

Hydrogen peroxide gas plasma works under deep vacuum. Radio frequency or microwave energy turns hydrogen peroxide into a plasma full of free radicals, which attack cell components.1 It runs at 37 to 44°C, has nontoxic by-products such as water vapor and oxygen, and needs no aeration. The cycle aborts if the load is wet. It cannot process paper, linens or liquids, so packs must use compatible wraps.1

Liquid chemical agents ("cold sterilization") such as glutaraldehyde are used for items with lenses, such as endoscopes and arthroscopes.2 Remember the time rule from the core concepts: a short soak is high-level disinfection, and only the long, label-specified soak reaches sterilization. Aldehydes irritate skin and mucous membranes, so items must be rinsed thoroughly with sterile saline before use.2 A cold tray on the counter that items sit in all day is a disinfection tray, not a sterile one.

Table 2. Sterilization and high-level disinfection methods compared
MethodHow it killsSuitable itemsCycle and aerationBiological indicator
Steam, gravity displacementCoagulates and denatures proteinsMetal instruments, cloth packs, heat-stable items121°C (250°F) for 30 min, wrapped; dry before removalG. stearothermophilus
Steam, prevacuumSameSame, including dense porous packs132°C (270°F) for 4 min, wrapped; daily Bowie-Dick testG. stearothermophilus
Ethylene oxideAlkylation of protein, DNA, RNAHeat- and moisture-sensitive plastics, rubber, scopesAbout 2.5 h; aerate 8 to 12 h at 50 to 60°CB. atrophaeus
Hydrogen peroxide gas plasmaFree radicalsHeat- and moisture-sensitive items; no paper, linen or liquid28 to 75 min by model; no aerationG. stearothermophilus
Liquid chemical (e.g. glutaraldehyde)AlkylationLensed instruments, semicritical itemsShort soak is high-level disinfection; long soak per label to sterilize; rinseNone routine

Sources: CDC Guideline for Disinfection and Sterilization;1 University of Guelph Veterinary Surgery Online.2

Monitoring: mechanical, chemical and biological

Three kinds of monitor work together. Mechanical (physical) monitors are the time, temperature and pressure readouts or printouts for each cycle. Chemical indicators change color when exposed to conditions in the cycle. Biological indicators (BIs) carry living spores, and only they measure the actual killing power of the process.1

The CDC guideline lists five classes of chemical indicator defined by the Association for the Advancement of Medical Instrumentation, in line with the ISO 11140-1 standard:15

  • Class 1, process indicator: autoclave tape on the outside of the pack. It only separates processed from unprocessed packs.5
  • Class 2, Bowie-Dick test: checks air removal in a prevacuum sterilizer.
  • Class 3, single-parameter and Class 4, multi-parameter indicators.
  • Class 5, integrating indicator: reacts to time, temperature and pressure together, so it tells you most about conditions at the center of the pack.5

Many steam strips used in practice change from white to black once a set temperature is reached. They show that the temperature was reached at some point, not how long it lasted.3 Plasma indicators usually change from red to yellow.3

Ladder of sterilization indicators from Class 1 tape to the biological indicator More information about the cycleBiological indicator (spore test)Living spores: the only test of killing powerClass 5 integrating indicatorInside pack: reacts to all critical variablesClass 4 multi-parameter indicatorInside pack: two or more variablesClass 3 single-parameter indicatorOne variable, such as temperatureClass 2 Bowie-Dick testPrevacuum only: air removal, empty chamberClass 1 process indicator (tape)Outside pack: processed or not, not sterilePhysical monitors (time, temperature and pressure readouts) are checked every cycle as well.
Figure 3. The indicator ladder. Each step up tells you more about whether the cycle could kill. Tape on the outside (Class 1) only shows the pack went through. An integrator inside the pack (Class 5) reacts to time, temperature and pressure. Only the biological indicator tests actual killing power.

Spore tests: G. stearothermophilus (105 spores) is incubated at 55 to 60°C. B. atrophaeus (106 spores) is incubated at 35 to 37°C.1 Steam and low-temperature sterilizers should be tested at least weekly. A busy sterilizer running several loads a day benefits from daily testing, which finds failures sooner.1 Results take time, from 1 to 7 days with conventional culture.3

Procedure: running a steam cycle

  1. Clean and dry. Remove all blood and debris. An ultrasonic cleaner helps clear box locks and serrations. Dry the instruments.2
  2. Open and arrange. Open box locks so every surface is exposed to steam.2
  3. Add an internal indicator. Place a chemical indicator inside each pack.1
  4. Wrap and label. Double wrap cloth or paper. Seal with indicator tape and write the date, contents and your initials on it.43
  5. Load loosely. Stand packs on edge with space between them. Place perforated trays parallel to the shelf, put basins and other solid containers on edge, place small items loosely in wire baskets, and stand peel pouches on edge in mesh racks.12
  6. Add the BI when scheduled. Follow the BI manufacturer's directions for placement, and incubate a positive control: a BI from the same lot that was not run through the cycle.1
  7. Run and record. Select the cycle in the manufacturer's instructions. Record the date, load contents, cycle time, temperature and pressure.1
  8. Dry and cool. Let the drying phase finish. Cool steam-sterilized items before handling them, and use heat-protective gloves for hot trays.1
  9. Check before storage. Confirm the tape changed and the pack is dry and intact. If the mechanical or chemical indicators suggest a failed cycle, do not use the load.1

Autoclave maintenance and daily checks

  • A prevacuum unit runs a Bowie-Dick test each day it is used: a test pack in an empty chamber, at the front bottom shelf over the drain, run at 134°C for 3.5 minutes. A uniform color change passes. A spot of unchanged color means trapped air.1
  • If the Bowie-Dick test fails, do not use the sterilizer until it has been inspected and passes the test.1
  • Keep maintenance records with dates for preventive maintenance and repairs, alongside the BI log.1
  • Follow the manufacturer for chamber cleaning, water quality and gasket care.

Storage and event-related shelf life

Store sterile packs in a limited-access, well-ventilated area protected from dust, moisture, insects and temperature and humidity extremes. Relative humidity in sterile storage should not exceed 70%.1 Do not crush, bend or puncture packs on the shelf.

Many facilities now use event-related shelf life: a pack remains sterile until an event contaminates it, such as a tear, wetting or a broken seal. Time-related practice puts an expiration date on each pack instead.1 Under either system, inspect the pack before use. If it is wet, torn or punctured, repack and reprocess it.1 Some practices still use time-based dates by wrap type. The Guelph text, for example, gives 4 weeks for double cloth and up to 1 year for paper-plastic and plastic pouches.4 Follow your hospital's written policy.

Flowchart: which method for this item?

Which sterilization or disinfection method for this item: a yes or no decision tree Start: clean the item firstRemove all visible soilQ1. Does it touch onlyintact skin?YesNoncritical: clean, thenlow-level disinfectantNoQ2. Can it take heat,moisture and pressure?YesSteam autoclave:wrap, load, run cycleNoQ3. Heat sensitive andtouches mucous membranes only?YesHigh-level disinfectionallowed (e.g. endoscope)NoQ4. Contains paper, linenor liquid?YesEthylene oxide,then full aerationNoLow-temperature methodper device labelGas plasma orethylene oxideEvery method: use indicators, label the pack,store it dry and closed, check before opening
Figure 1. Choosing a reprocessing method. Always clean first. Items that touch only intact skin need low-level disinfection. Heat- and moisture-stable items go in the steam autoclave. Heat-sensitive semicritical items may receive high-level disinfection. Heat-sensitive critical items go to a low-temperature sterilizer: ethylene oxide if they contain paper, linen or liquid, otherwise gas plasma or EO as the device label allows.

Species differences

The method depends on the item, not the species. Dog and cat surgical packs, feline spay packs and equine or bovine packs all follow the same Spaulding logic, indicators and storage rules.

  • Dentistry (dogs and cats): instruments that cut soft tissue or bone, such as extraction forceps, scalpel blades, periodontal scalers and surgical burs, are critical and should be sterilized after each use or discarded.1
  • Large animal and field surgery: packs for horses and cattle are often bigger and heavier. Large, dense packs slow steam penetration, so loading spacing and an internal indicator at the center matter even more. A cold tray carried in a truck gives disinfection, not sterility.
  • Exotics and endoscopy: small rigid scopes and fine instruments used in birds, reptiles and small mammals are often heat sensitive. Check the device label for steam, EO or gas plasma compatibility before choosing.

Calculation: converting cycle temperatures

Autoclave panels may display Celsius or Fahrenheit. Use °F = (°C × 9/5) + 32.

Example: a gravity displacement cycle is set to 121°C. What is that in Fahrenheit?

121°C × 9/5 = 217.8, then 217.8 + 32 = 249.8°F, which rounds to 250°F.

Check: the prevacuum setting of 132°C gives 132 × 9/5 = 237.6, plus 32 = 269.6°F, about 270°F. Both match the CDC values of 250°F and 270°F.1

  • Ethylene oxide should be treated as a known human carcinogen. The OSHA permissible exposure limit is 1.0 ppm as an 8-hour time-weighted average.1 Never cut aeration short to get an item back faster.
  • Burns: cool instruments before use. Hot instruments can burn staff and patients; air-cool them or immerse in sterile saline.1
  • Aldehydes irritate skin, eyes and airways and can cause hypersensitivity. Wear gloves, keep containers covered and rinse items fully.2
  • Immediate-use (flash) steam sterilization is not for convenience or to avoid buying more instruments, and it is not recommended for implants. If an implant must be flashed, record the load, the patient and the BI result.1
  • Records: keep cycle logs, BI results and maintenance records. They show the practice can trace a failed load back to the patients who received items from it.1
  • Client education: clients may ask how instruments are kept safe. A simple answer is that every pack is cleaned, run through a monitored cycle and checked with indicators, and the sterilizer is spore tested on a schedule.

Common mistakes and exam traps

  • "The tape changed, so it is sterile." Wrong. External tape is a Class 1 process indicator. Only the BI tests killing power.1
  • Pressure kills microbes. No. Pressure lets steam reach a higher temperature; moist heat kills.1
  • Cold sterile tray equals sterile. A short soak is high-level disinfection. Spores survive.1
  • Wrong spore for the method. G. stearothermophilus for steam and gas plasma; B. atrophaeus for EO. Older texts name the EO organism Bacillus subtilis; B. atrophaeus is the current name.13
  • Linen in the plasma sterilizer. Gas plasma cannot process paper, linens or liquids.1
  • Packing the chamber tight. Overloading traps air and leaves cold spots. Load loosely, on edge.1

Quick review

  1. Clean before any disinfection or sterilization.
  2. Critical items must be sterile; semicritical need at least high-level disinfection.
  3. Gravity displacement: 121°C (250°F) for 30 minutes, wrapped.
  4. Prevacuum: 132°C (270°F) for 4 minutes, wrapped; Bowie-Dick test daily.
  5. Moist heat kills by coagulating proteins; pressure only raises the temperature.
  6. Ethylene oxide needs aeration, 8 to 12 hours at 50 to 60°C by machine.
  7. Gas plasma: no paper, linen or liquid; no aeration needed.
  8. Short glutaraldehyde soak is high-level disinfection, not sterilization.
  9. Class 1 tape shows processing only; Class 5 integrator reacts to time, temperature and pressure.
  10. Spore test at least weekly: G. stearothermophilus for steam, B. atrophaeus for EO.
  11. Load loosely, on edge; cool before handling.
  12. Event-related shelf life: wet, torn or punctured means reprocess.

Test yourself

5 questions written for this guide. Pick an answer, then check it.

Question 1On Monday morning a technician runs the Bowie-Dick test in the hospital's empty prevacuum autoclave. The test sheet shows a pale, unchanged spot in the center while the edges changed color. What should the technician do?

Show answer

Answer: C. Take the sterilizer out of use until it is inspected and passes a repeat Bowie-Dick test

A spot of unchanged color means air was trapped and steam did not reach the center of the test pack. CDC guidance says a sterilizer that fails the Bowie-Dick test should not be used until it is inspected and passes the test. A is wrong: a passing sheet shows a uniform color change. B is wrong: integrators do not fix an air-removal fault in the machine. D is wrong: 15 minutes is below the recognized 30 minutes for wrapped packs in a gravity cycle, and the fault is still unrepaired.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Steam Sterilization, CDC

Question 2A heat-sensitive rubber and plastic device must be sterilized, and the only wrap available in the hospital is linen. Which method is most appropriate?

Show answer

Answer: A. Ethylene oxide sterilization, followed by a full aeration period before use

Ethylene oxide works at low temperature and can be used with cloth wraps, but absorbent materials hold toxic residue, so aeration is required. B is wrong: gas plasma cannot process linens or paper. C is wrong: steam would damage a heat-sensitive item. D is wrong: a 20-minute glutaraldehyde soak is high-level disinfection, not sterilization.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Ethylene Oxide and Hydrogen Peroxide Gas Plasma, CDC

Question 3A practice buys a hydrogen peroxide gas plasma sterilizer. Which biological indicator and incubation temperature should the technician use to monitor it?

Show answer

Answer: D. Geobacillus stearothermophilus spores, incubated at 55 to 60°C

CDC lists G. stearothermophilus spores for steam, hydrogen peroxide gas plasma and liquid peracetic acid, incubated at 55 to 60°C. A is the ethylene oxide and dry heat indicator. B pairs the EO organism with the wrong temperature. C pairs the right organism with the B. atrophaeus incubation temperature.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Monitoring (biological indicators), CDC

Question 4An autoclave panel displays only Fahrenheit. The prevacuum cycle should run at 132°C. Using °F = (°C x 9/5) + 32, which setting is closest?

Show answer

Answer: B. 270°F

132 x 9/5 = 237.6, and 237.6 + 32 = 269.6°F, about 270°F, the CDC prevacuum value. A is 121°C, the gravity displacement temperature. C forgets to add 32. D overshoots the conversion.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Steam Sterilization, CDC

Question 5A hospital uses event-related shelf life. A double-wrapped pack has been on a closed shelf for 9 months. The wrap is dry and intact and the tape has changed color. What is the best action?

Show answer

Answer: C. Inspect it and use it, because no event has compromised the package

Under event-related practice a pack stays sterile until an event such as wetting, tearing, puncture or a broken seal contaminates it. The technician checks integrity before use. A and B apply time-based dates that the policy does not use. D is wrong: a spore test checks the sterilizer cycle, not stored packs, and opening the pack ends its sterility.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Storage of Sterile Items, CDC

Practise this topic

Two questions from our VTNE practice bank on this topic.

Question 1A veterinary technician reviews the quality assurance records for the autoclave. What is the fundamental difference between a chemical indicator and a biological indicator?

Show answer

Answer: A. Chemical ones sense conditions; biological ones prove spore kill

Chemical indicators are heat- or chemical-sensitive inks that change color when sterilization conditions such as temperature or steam are reached. They show that a pack was processed, but not how long conditions were held, so they do not prove sterility. Biological indicators contain highly resistant bacterial spores, such as Geobacillus (Bacillus) stearothermophilus for steam, and are the only process indicators that directly measure whether the cycle killed microorganisms. Results take about 1 to 7 days.

The description of biological indicators as color strips and chemical indicators as bacteria to incubate reverses the two. They do not work the same way: one measures physical conditions and the other measures actual microbial kill. Chemical indicators are not the gold standard and need no incubation; biological indicators are the ones that require culture and are considered the most accurate monitor.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), US Centers for Disease Control and Prevention
Sterilization Indicators, Veterinary Surgery Online

Question 2A veterinary practice uses event-related rather than time-related sterility policies for wrapped surgical packs. Under an event-related policy, what PRIMARILY determines when a pack is no longer sterile?

Show answer

Answer: A. An event such as a tear, moisture or dropping that damages the package

Event-related shelf life assumes a properly stored pack stays sterile until something happens to compromise it, such as a tear, a puncture, a broken seal or the wrap getting wet. Handling, traffic, humidity, pests and the type of wrap all affect the risk, and contamination becomes more likely the more a pack is handled. Under this policy packs can be used indefinitely unless the packaging is compromised, so every pack is inspected for integrity before it is opened.

Fixed limits such as 30 days or 1 year belong to time-related policies, where each pack is dated and reprocessed after a set shelf life, not to event-related policies. Autoclave tape only shows that a certain temperature was reached during the cycle; its color is not used to judge how long a pack stays sterile in storage.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008, US Centers for Disease Control and Prevention
Sterilization Indicators, Veterinary Surgery Online

Figure credits

  • Figure 1 Original diagram, VTNE Exam Prep
  • Figure 2 Original diagram, VTNE Exam Prep
  • Figure 3 Original diagram, VTNE Exam Prep

References

  1. Rutala WA, Weber DJ, Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008 (last update June 2024). Centers for Disease Control and Prevention. cdc.gov (accessed 2026-10-31) cdc.gov/infection-control/media/pdfs/guideline-disinfection-h.pdf
  2. University of Guelph, Ontario Veterinary College. Veterinary Surgery Online, Chapter 3: Sterilization techniques. vetsurgeryonline.com (accessed 2026-10-31) vetsurgeryonline.com/sterilization-techniques/
  3. University of Guelph, Ontario Veterinary College. Veterinary Surgery Online, Chapter 3: Sterilization Indicators. vetsurgeryonline.com (accessed 2026-10-31) vetsurgeryonline.com/sterilization-indicators/
  4. University of Guelph, Ontario Veterinary College. Veterinary Surgery Online, Chapter 3: Sterilization packaging. vetsurgeryonline.com (accessed 2026-10-31) vetsurgeryonline.com/sterilization-packaging/
  5. Kumaran M, Nirmal F, Gagan D, Sukumaran R, Chandrasekar D. Steam sterilization: review of autoclaves, validation, instrument packing, and sterility failures. Indian J Ophthalmol. 2026;74(2):173-182. PMC12947077 pmc.ncbi.nlm.nih.gov/articles/PMC12947077/
  6. American Association of Veterinary State Boards. Veterinary Technician National Examination (VTNE) Content Outline (2023). aavsb.org (accessed 2026-10-31) aavsb.org/wp-content/uploads/2025/04/AAVSB-2023-Veterinary-Technician-National-Examination-Content-Outline-2.pdf

Spotted an error? Email support@vtneexam.com and we will check it against the sources.