Diagnostic Imaging Subscriber guide

Ultrasound, Endoscopy, CT and MRI: The Technician's Role

Diagnostic Imaging (D6 A, B, C; K39, K40, K45) 11 min read, 2,413 words Last verified

How to choose a transducer and read echogenicity terms, prepare patients for CT, MRI and endoscopy, keep the MRI suite safe, and reprocess flexible endoscopes.

Introduction

Radiography is not the only imaging tool in a modern practice. Ultrasound is probably the second most commonly used imaging method in veterinary practice, and many referral and some general practices also run computed tomography (CT) and magnetic resonance imaging (MRI) units.13 Flexible endoscopes let the veterinarian look inside the airway, esophagus, stomach and colon.

The veterinarian or radiologist reads the images. The credentialed veterinary technician prepares and positions the patient, picks and sets up the equipment, runs anesthesia, keeps everyone safe around the magnet, and cleans and stores delicate scopes. The VTNE places this material in Domain 6, Diagnostic Imaging, and in the equipment care knowledge area.10

Key facts

  • Diagnostic ultrasound uses sound waves, most often 1.5 to 18 megahertz (MHz). A transmission gel couples the transducer to the skin.1
  • Low-frequency transducers scan deeper, but resolution is lower. High frequency gives finer detail at shallow depth.1
  • Sound is almost totally reflected at soft tissue and gas interfaces and absorbed at bone, so gas and bone cast shadows.1
  • CT measures density in Hounsfield units (HU): air is -1,000, water is 0. Nearly all veterinary soft tissue CT studies use contrast.3
  • MRI uses no ionizing radiation, but its magnet can pull an oxygen cylinder into the scanner at nearly 100 miles per hour. Small animal MRI is done under general anesthesia.2
  • An MR suite has four zones. Zone IV is the magnet room and carries the highest risk.4
  • Leak test a flexible endoscope before immersion, then clean, high-level disinfect, rinse, dry and store it hanging vertically.6

How the VTNE tests this

  • Choose the transducer: match frequency to patient size and the depth of the organ (Domain 6 Task B; K39).10
  • Use echogenicity terms: name an anechoic, hypoechoic or hyperechoic structure, or an acoustic shadow (Domain 6 Task A; K39).
  • Prepare the patient: clipping, gel, fasting and enemas for endoscopy, anesthesia for CT and MRI (Domain 6 Task B).
  • Apply MRI safety: screen people, patients and equipment for ferromagnetic material before Zone IV (Domain 6 Task B; K40).
  • Care for equipment: endoscope leak testing, reprocessing order and storage (Domain 6 Task C; K45).10

Core concepts

Ultrasound physics in plain words

The transducer (probe) sends a very short pulse of sound into the patient, then switches to listen. An echo forms wherever the sound passes from tissue of one density to another. The bigger the change, the stronger the echo. The machine records how strong each echo is and how long it took to return, and builds a real-time grayscale picture called B-mode.1 Images are saved in DICOM format, the accepted legal format, just like digital radiographs.1

Frequency versus penetration. This is the most tested physics rule. Low-frequency transducers scan deeper than high-frequency transducers, but resolution is decreased.1 A WSAVA congress lecture gives typical choices: about 5.0 MHz for large patients, and 7.5 or 10 MHz for small to medium dogs, cats and superficial structures.8 In cats, the frequency is often set to the highest level because the patient is small.9 The operator also adjusts depth and gain (overall brightness).9

Transducer frequency compared with depth and detail Frequency trades depth for detail Skin Deep Low (about 5 MHz) Reaches deepest Coarser detail Medium (7.5 MHz) Middle depth Small and medium dogs High (10 MHz +) Shallow only Finest detail Cats, superficial
Figure 1. Bar height shows how deep each frequency band can image. Lower frequency reaches deeper tissue but with less detail; higher frequency gives finer detail only near the skin. Bar heights are schematic, not to scale.

Echogenicity terms

Echogenicity is how bright a tissue looks, and it is always judged by comparing one tissue with another.18 Learn the words below. Fluid such as urine or bile is the classic anechoic structure.

Table 1. Ultrasound echogenicity and artifact terms
TermWhat you seeTypical example
AnechoicBlack, no echoesFluid, such as pleural effusion or urine
HypoechoicDarker than surrounding tissueNormal liver compared with spleen
HyperechoicBrighter than the comparison tissueFat; reverberation lines from gas
Acoustic shadowAnechoic band deep to a strong reflector or absorberBone, gas, uroliths, gallstones
Acoustic enhancementBright zone deep to a fluid-filled structureDeep to the gallbladder
ReverberationEvenly spaced bright lines deep to a strong reflectorGas; A-lines from the lung surface

Sources: Swanstein and colleagues, JFMS 2024; Boroffka, WSAVA 2015; Armenise, Animals 2025; Merck Veterinary Manual, Feline Hepatic Lipidosis.981112

CT in plain words

CT passes an x-ray beam through the patient from many angles. A computer works out the density of each small volume of tissue (a voxel) and builds a cross-section called a slice. Density is given in Hounsfield units, from -1,000 for air through 0 for water to +3,000 for heavy metal.3 Because slices do not overlap, CT shows structures without superimposition and has much better contrast resolution than radiographs.3 CT uses ionizing radiation, so the radiation safety rules from radiography still apply.

MRI in plain words

MRI uses a powerful magnet, up to 60,000 times the strength of the earth's field, to line up hydrogen atoms. Radiofrequency pulses tip them, and as they relax they give off signals that the scanner turns into images.2 Each sequence can take several minutes, and MRI is very sensitive to motion, so small animal studies are done under general anesthesia.2 MRI shows soft tissue, cartilage and the nervous system in great detail but images cortical bone less well than CT.2

The magnet is the hazard. Ferromagnetic material (iron and most steel objects) cannot be brought into the room, because the field can accelerate an object such as an oxygen bottle to nearly 100 miles per hour.2 The FDA lists projectile events, burns, crushed fingers and hearing loss among reported MRI injuries.5 Metal also ruins images: even the iron in an identity microchip, or a single steel BB pellet, can degrade the study, so radiographs of the area of interest are taken before MRI.2

Table 2. The four MR safety zones (American College of Radiology model)
ZoneAreaAccess
IPublic areas outside the MR environment, such as the waiting roomUnrestricted
IIArea between I and III where patients are brought in and screenedSupervised
IIIControl roomHighly restricted; MR personnel only
IVScanner magnet room, highest riskTrained personnel; door locked when unattended

Source: Shah and Aran, Cureus 2023, summarizing the ACR Manual on MR Safety.4

Procedure or technique

Abdominal ultrasound: the technician's steps

  1. Confirm the request and the patient's identity. Ask the veterinarian whether the patient should be fasted.
  2. Restrain the patient in dorsal recumbency. Abdominal ultrasound is usually done in the conscious animal.8
  3. Clip the hair over the ventral abdomen and apply contact gel for good transducer to skin contact.8 Air between probe and skin blocks the sound.1
  4. Select the transducer and preset for the patient's size, then adjust depth and gain.9
  5. Label and save images in DICOM format to the patient record.1
  6. For ultrasound-guided sampling, set up the needle, syringe and slides. Guidance is safer and more diagnostic than a blind biopsy.1
  7. Wipe gel off the patient and clean the transducer as the manufacturer directs.
A military working dog lies on its back on a padded table while a radiologist holds an ultrasound transducer on its clipped abdomen and watches the screen
Figure 3. Abdominal ultrasound of a dog in dorsal recumbency. The hair is clipped and the transducer is held on the ventral abdomen while the operator watches the real-time image. Lt. Cmdr. Jesse Ehrenfeld, U.S. Navy, public domain.

CT: preparation and positioning

  1. Complete the pre-anesthetic workup. Most veterinary CT studies need anesthesia and immobilization, though the anesthetic period is short.3
  2. Place an IV catheter for contrast. Nearly all veterinary CT studies of soft tissue, such as the brain and abdomen, are contrast enhanced, and nonionic contrast agents improve diagnostic accuracy.3
  3. Position the patient straight and symmetrical with foam pads and ties. Scans of awake or lightly sedated animals lack controlled positioning, show movement and are hard to interpret.3
  4. Monitor the patient through the scan and after the contrast injection, and record the contrast agent and volume.

MRI: preparation

  1. Screen the patient: microchip, implants, surgical clips, gunshot. Radiograph the area of interest first.2
  2. Screen yourself and every handler. Remove keys, phones, pens, scissors, stethoscopes and clipboards. Use MR-labeled equipment only; ASTM F2503 labels devices MR Safe, MR Conditional or MR Unsafe.4
  3. Induce anesthesia outside Zone IV. Use MR-compatible anesthesia machines, oxygen supply and monitors; if these are not available, injectable anesthesia can be used.2
  4. Provide hearing protection. Gradient noise can reach 100 dB or more.4
  5. In an emergency, MR-trained staff move the patient out of Zone IV to a magnetically safe area before resuscitation.4
A veterinary health technician checks the pulse of an anesthetized dog lying on the table of an MRI scanner
Figure 4. A veterinary health technician takes the pulse of an anesthetized military working dog during an MRI exam. The patient stays anesthetized and monitored for the whole study. TSgt Lance Cheung, USAF, 1994, public domain.

Endoscopy: patient prep and scope care

Endoscopy is done under general anesthesia to protect both the patient and the expensive scope.7 Fast most patients 12 to 24 hours before upper GI endoscopy, longer (24 to 48 hours) if gastric emptying is delayed. For colonoscopy, fast 24 to 48 hours and give warm-water enemas the evening before and again 2 to 4 hours before, until the return is clear.7 Do not use Fleet enemas (a mildly hypertonic enema solution) in cats or small dogs.7

Flexible endoscopes are heat sensitive, semicritical items (they touch mucous membranes). They must be cleaned and, at a minimum, high-level disinfected after each use.6 Meticulous cleaning must come before disinfection.6

Table 3. Flexible endoscope reprocessing order
StepWhat the technician does
Leak testBefore immersion, check the scope for leaks so fluid cannot enter a damaged scope
1. CleanBrush internal channels; flush each channel with water and detergent or enzymatic cleaner
2. DisinfectImmerse in high-level disinfectant and perfuse all channels for the product's contact time
3. RinseRinse the scope and all channels with sterile, filtered or potable tap water
4. DryFlush with alcohol and dry with forced air
5. StoreHang vertically in a way that prevents recontamination and promotes drying

Source: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities.6

Table 4. Modality, preparation and safety at a glance
ModalityPatient prepMain safety point
UltrasoundClip, gel; often awake with restraintNo ionizing radiation; gel and probe hygiene
CTAnesthesia, IV catheter, contrastIonizing radiation; contrast monitoring
MRIAnesthesia, metal screening, radiographs firstProjectiles in Zone IV; burns; noise
EndoscopyAnesthesia; fasting; enemas for colonLeak test and high-level disinfection

Sources: Merck Veterinary Manual; Ford and Mazzaferro; CDC; Shah and Aran.123467

Flowchart

MRI screening checklist before entering Zone IV Patient booked for MRI Microchip, implant, clips or gunshot in the history or on radiographs? Yes Tell the veterinarian; wait for a plan No Every person screened? Keys, phone, pens, scissors, stethoscope removed? No Stop in Zone II or III Yes All equipment labeled MR Safe or MR Conditional for this scanner? No Swap for MR equipment Yes Patient induced outside Zone IV, hearing protection in place? No Finish prep outside Yes Enter Zone IV with trained staff Emergency in the scanner: move the patient out of Zone IV, then start resuscitation. Keep steel equipment out of Zone IV.
Figure 2. MRI screening checklist. Every "no" stops the team outside the magnet room until the problem is fixed. Based on Merck Veterinary Manual and the ACR zone model summarized by Shah and Aran.

Species differences

  • Dogs: transducer choice depends on size. A large, deep-chested dog may need about 5 MHz; small and medium dogs use 7.5 or 10 MHz.8
  • Cats: small size means the highest frequency setting is often used, and presets work well because cats vary less in size than dogs.9 Clip only what is needed and keep the cat warm.
  • Horses: ultrasound is used most often for tendon and ligament tears, and transrectal ultrasound for reproduction.1 Standing MRI of the distal limb is done under sedation; recumbent MRI needs general anesthesia.2 Standing CT systems with a moving gantry exist for sedated horses.3
  • Ruminants and other large animals: transrectal ultrasound of the reproductive tract is routine, and ultrasound-guided sampling can be done without general anesthesia.1

Safety, legal and client education

  • MRI uses no ionizing radiation, which protects staff.2 CT uses an x-ray beam, so follow the same radiation safety rules as for radiography.3
  • Lock the Zone IV door when no one is attending it, and keep it under supervision by trained personnel.4
  • A quench is the rapid release of the magnet's liquid helium through a vent pipe. If the pipe fails, helium in the room can lower oxygen levels.4
  • Saved ultrasound, CT and MRI images are part of the medical record; the DICOM format is the legal standard.1
  • Explain to clients that clipping is needed for a clear ultrasound image, that CT and MRI need anesthesia because any movement ruins the study, and why their pet must be fasted before endoscopy.237
  • Endoscope reprocessing staff should get initial and annual competency testing.6

Common mistakes and exam traps

  • Reversing the frequency rule. Higher frequency means better detail but less depth, not more depth.1
  • Calling fluid hyperechoic. Simple fluid is anechoic (black); bone and gas cause shadows.19
  • Thinking a microchip is harmless in MRI. Its iron can cause significant image artifacts.2
  • Running into Zone IV with a steel oxygen cylinder or crash cart during an emergency. Move the patient out instead.24
  • Soaking an endoscope before the leak test, or storing it coiled and wet. Leak test first, dry fully, hang vertically.6
  • Assuming CT is done awake. Most veterinary CT studies need anesthesia.3

Memory aids

LOW goes LOW: low frequency goes low (deep) in the body; high frequency stays high near the skin with finer detail. (memory aid)

HYPER is HIGHLIGHTED: hyperechoic structures are brighter; hypo is darker; anechoic means "no echo" and looks black. (memory aid)

Zone IV, the fourth wall is the magnet: zones count up from the waiting room (I) to the magnet room (IV), where the risk is highest. (memory aid)

Leak before you soak: leak test a flexible endoscope before immersing it. (memory aid)

Can Doctors Really Dry Scopes: Clean, Disinfect, Rinse, Dry, Store, the CDC endoscope reprocessing order. (memory aid)

Quick review

  1. Diagnostic ultrasound uses about 1.5 to 18 MHz; gel couples probe to skin.
  2. Low frequency penetrates deeper; high frequency gives better resolution.
  3. About 5 MHz for large patients; 7.5 to 10 MHz for small dogs and cats.
  4. Anechoic is black (fluid); hypoechoic is darker; hyperechoic is brighter.
  5. Gas reflects and bone absorbs sound, leaving acoustic shadows.
  6. Abdominal ultrasound: dorsal recumbency, clip, gel, often awake.
  7. CT density is in Hounsfield units: air -1,000, water 0.
  8. Most veterinary CT and all small animal MRI studies need anesthesia.
  9. Radiograph the area before MRI; microchips and BB pellets cause artifacts.
  10. MR zones I to IV; Zone IV is the magnet room with the highest risk.
  11. Upper GI endoscopy: fast 12 to 24 hours; colonoscopy: 24 to 48 hours plus enemas.
  12. Endoscope care: leak test, clean, disinfect, rinse, dry, hang vertically.

Test yourself

5 questions written for this guide. Pick an answer, then check it.

Question 1A technician is setting up for an abdominal ultrasound on a 42 kg deep-chested dog. The veterinarian cannot see the far side of the liver with a 10 MHz transducer. Which change will most likely help?

Show answer

Answer: C. Switch to a lower-frequency transducer of about 5 MHz

Lower-frequency transducers penetrate deeper, at the cost of resolution, and about 5 MHz is the usual choice for large patients. A makes penetration worse. B brightens the image but does not add depth. D improves contact only; it cannot overcome the depth limit of a high frequency.

Source Ultrasonography in Animals, Merck Veterinary Manual

Question 2On an ultrasound image, urine inside the bladder appears completely black. Which term describes this?

Show answer

Answer: A. Anechoic

Anechoic means no echoes, which shows as black; simple fluid is the classic example. B means brighter than the comparison tissue. C means the same brightness as the comparison tissue. D is a CT density term, not an ultrasound term.

Source Feline friendly POCUS: how to implement it into your daily practice, J Feline Med Surg 2024

Question 3A dog is booked for brain MRI. During screening, the owner reports the dog has a microchip and was once shot with a BB gun. What should the technician do first?

Show answer

Answer: B. Report the history so radiographs of the area can be taken

Merck advises radiographs of the area of interest before MRI, because even microchip iron or a single pellet can degrade the images; the veterinarian decides the plan. A is not routine. C is false: both can cause artifacts. D is a veterinary decision, not the technician's.

Source Magnetic Resonance Imaging in Animals, Merck Veterinary Manual

Question 4Which step must come first when reprocessing a flexible gastroscope after use?

Show answer

Answer: D. Leak testing it before it is immersed

CDC lists the reprocessing steps after leak testing, and leak testing is done before immersion so fluid cannot enter a damaged scope. A is step 2, after cleaning. B is the drying step. C is storage, the last step.

Source Guideline for Disinfection and Sterilization in Healthcare Facilities, CDC

Question 5In the four-zone MR safety model, where is the scanner magnet located?

Show answer

Answer: C. Zone IV, the magnet room itself

The magnet is in Zone IV, the most restricted, highest-risk area, locked when unattended. Zone I is public, Zone II is for screening and Zone III is the control room.

Source A Review of Magnetic Resonance (MR) Safety: The Essentials to Patient Safety, Cureus 2023

Practise this topic

22 questions from the practice bank are linked to this guide. Answer them in practice mode with instant explanations.

Figure credits

References

  1. Boyle AG. Ultrasonography in Animals. Merck Veterinary Manual. merckvetmanual.com, last updated April 2025 (accessed 2026-11-04) merckvetmanual.com/clinical-pathology-and-procedures/diagnostic-imaging/ultrasonography-in-animals
  2. Manzi T. Magnetic Resonance Imaging in Animals. Merck Veterinary Manual. merckvetmanual.com (accessed 2026-11-04) merckvetmanual.com/clinical-pathology-and-procedures/diagnostic-imaging/magnetic-resonance-imaging-in-animals
  3. Manzi T. Computed Tomography in Animals. Merck Veterinary Manual. merckvetmanual.com (accessed 2026-11-04) merckvetmanual.com/clinical-pathology-and-procedures/diagnostic-imaging/computed-tomography-in-animals
  4. Shah A, Aran S. A Review of Magnetic Resonance (MR) Safety: The Essentials to Patient Safety. Cureus. 2023;15(10):e47345. PMC10657250 pmc.ncbi.nlm.nih.gov/articles/PMC10657250/
  5. US Food and Drug Administration. MRI (Magnetic Resonance Imaging): Benefits and Risks. fda.gov (accessed 2026-11-04) fda.gov/radiation-emitting-products/mri-magnetic-resonance-imaging/benefits-and-risks
  6. Rutala WA, Weber DJ, Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008 (updated). CDC. cdc.gov (accessed 2026-11-04) cdc.gov/infection-control/media/pdfs/guideline-disinfection-h.pdf
  7. Ford RB, Mazzaferro EM. Diagnostic and Therapeutic Procedures. In: Kirk and Bistner's Handbook of Veterinary Procedures and Emergency Treatment. 9th ed. Elsevier; 2012. PMC7158366 pmc.ncbi.nlm.nih.gov/articles/PMC7158366/
  8. Boroffka S. An Ultrasonic Rollercoaster Through the Canine and Feline Abdomen. WSAVA World Congress Proceedings, 2015. vin.com (accessed 2026-11-04) vin.com/apputil/content/defaultadv1.aspx?id=7259379&pid=14365&print=1
  9. Swanstein H, Boysen S, Cole L. Feline friendly POCUS: how to implement it into your daily practice. J Feline Med Surg. 2024;26(9):1098612X241276916. PMC11418624 pmc.ncbi.nlm.nih.gov/articles/PMC11418624/
  10. American Association of Veterinary State Boards. Veterinary Technician National Examination (VTNE) Content Outline, approved March 2023. aavsb.org (accessed 2026-11-04) aavsb.org/wp-content/uploads/2025/04/AAVSB-2023-Veterinary-Technician-National-Examination-Content-Outline-2.pdf
  11. Armenise A. Point-of-Care Lung Ultrasound in Small Animal Emergency and Critical Care Medicine: A Clinical Review. Animals (Basel). 2025. PMC11718948 pmc.ncbi.nlm.nih.gov/articles/PMC11718948/
  12. Merck Veterinary Manual. Feline Hepatic Lipidosis. merckvetmanual.com (accessed 2026-11-04) merckvetmanual.com/digestive-system/hepatic-diseases-of-small-animals/feline-hepatic-lipidosis

Spotted an error? Email support@vtneexam.com and we will check it against the sources.