Hemostasis and Electrosurgery: Pressure, Ligation, Energy Devices and Topical Agents

Surgical Nursing (D2 D, E; K19) 11 min read, 2,313 words Last verified

How surgical bleeding is controlled and what the technician does: pressure, ligatures and clips, electrosurgery and vessel sealers, lasers and smoke, topical agents and estimating blood loss by weighing sponges.

Introduction

Hemostasis means stopping bleeding. In surgery it is done with pressure, clamps, ligatures, clips, energy devices and topical agents. The surgeon picks the method, but the technician sets up and runs most of the equipment: you place the return electrode, set the electrosurgical unit, run the smoke evacuator, hand the right hemostat, open topical agents onto the field and weigh sponges to estimate blood loss. Mistakes in this work cause patient burns, surgical fires and staff exposure to smoke. This guide covers what each method does, how to run the equipment safely and how to calculate blood loss, which are all fair game on the VTNE.

Key facts

  • Blot a bleeding field with gauze; rubbing can dislodge small clots and restart bleeding.2
  • In monopolar electrosurgery, current flows from the active tip through the patient to a wide return electrode (ground plate) and back to the generator.1
  • Poor contact or partial detachment of the return electrode raises current density and can burn the patient under the pad.1
  • Bipolar devices use lower voltage and two electrodes close together, which almost removes the risk of alternate site burns.1
  • Advanced bipolar vessel sealers such as LigaSure have been shown to seal vessels up to 7 mm in diameter.4
  • Bone wax stops bleeding from cut bone mechanically, but it is poorly absorbed and can impair bone healing.23
  • The sponge weight (gravimetric) method assumes 1 g of blood equals 1 mL.5
  • A surgical fire needs an oxidizer, an ignition source and fuel; high oxygen concentration is the most important factor.1

How the VTNE tests this

  • Choose the method: match a bleeding situation (cut bone, oozing organ surface, large pedicle, small skin bleeder) to the hemostatic method or agent (Domain 2 Task D, sterile technician).11
  • Set up the equipment: where the ground plate goes, when bipolar is preferred, what the smoke evacuator is for (Domain 2 Task E, circulating technician; Knowledge 19, safety of surgical equipment).11
  • Recognize a hazard: a burn under the pad, an alcohol prep with an energy device, oxygen near a laser beam.
  • Calculate: blood loss from sponge weights and the suction canister, then compare it with the patient's blood volume.

Core concepts

Mechanical methods

Time and pressure. Small vessels clot on their own given time. Direct pressure helps, but too much pressure fully occludes the vessel and keeps platelets and clotting factors from reaching the site. Apply just enough pressure for long enough, and blot rather than rub.2

Hemostatic forceps (hemostats) both occlude a vessel and crush its wall, which stimulates clotting. A small vessel can often be released after a few minutes.2

Ligation means tying a vessel with suture. If a hemostat is holding the vessel, the surgeon asks for it to be "flashed" (opened briefly) as the knot is tightened so the ligature seats properly. A transfixation ligature passes through the stump before it is tied, so it is less likely to slip. It is common on the uterine stump in a spay and on large vessels in splenectomy or amputation.2

Vascular clips are quick. Choose a clip about one third to two thirds longer than the vessel diameter, because a flattened vessel becomes wider.2

Tourniquets reduce bleeding during distal limb surgery such as toe amputation, but bleeding may start when the tourniquet comes off, so the surgeon may release it before closing to check.2

Electrosurgery

Electrosurgery uses high-frequency electrical current to cut or coagulate tissue. It works well for small arteries up to about 1 mm and veins up to about 2 mm.2 Two circuit designs exist (Figure 1).

  • Monopolar: the current is concentrated at the small active tip, passes through the patient and returns to the generator through a wide return electrode with low current density.1 The return electrode is also called the dispersive electrode, ground plate or ground pad.
  • Bipolar: both electrodes sit in the forceps tips, so current passes only through the tissue grasped between them. Lower voltage and close electrodes almost remove the risk of alternate site burns.1 No ground plate is needed. Bipolar is chosen near delicate structures where current spread matters.
  • Advanced bipolar vessel sealers (LigaSure, EnSeal) clamp the tissue, seal it under device control and often divide it. LigaSure has been shown to seal vessels up to 7 mm in diameter and has been called a gold standard in laparoscopic surgery. LigaSure and EnSeal have been compared for laparoscopic ovariectomy in dogs.4

Lateral thermal spread is heat moving from the device into nearby tissue. It is the most common way electrosurgery injures tissue, and the risk rises with higher power and longer activation.1

Current path in monopolar and bipolar electrosurgery Monopolar Bipolar Generator Generator Active pencil Patient Patient Return electrode (ground plate) Forceps Current stays No ground plate needed between tips
Figure 1. In monopolar electrosurgery (left) current spreads from the small active tip through the patient to the large return electrode, so pad contact matters. In bipolar electrosurgery (right) current passes only through tissue held between the forceps tips. Schematic, not to scale. Original diagram, VTNE Exam Prep.

Topical hemostatic agents

Topical agents help with oozing that is hard to clamp or tie. They do not replace ligation of a named vessel.

Table 1. Hemostatic methods: mechanism, typical use and main risk
MethodMechanismTypical useMain risk or caution
Direct pressureSlows flow so a clot can formSmall bleeders, first responseToo much pressure blocks clot formation; rubbing restarts bleeding
Hemostatic forcepsOcclude and crush the vessel wallSmall to medium vesselsMust be flashed during ligation
Ligature or transfixationSuture occludes the vesselPedicles, large vesselsSlippage; transfixation reduces it
Vascular clipsMetal or polymer clip occludes the vesselQuick ligation, spay pediclesClip too short for the flattened vessel
Monopolar electrosurgeryCurrent through patient to ground plateSmall vessels, cuttingPad burns, fire, smoke
Bipolar or vessel sealerCurrent only between jawsNear nerves, vessels up to 7 mm (sealer)Lateral thermal spread
Bone waxPlugs channels in cut boneCut bone surfaces, e.g. mandibulectomyPoorly absorbed; impairs bone healing; infection
Gelatin sponge (Gelfoam, Vetspon)Absorbs fluid, swells, forms a matrixOozing surfacesNo antimicrobial effect
Oxidized regenerated cellulose (Surgicel)Acidic; procoagulantCapillary, venous, minor arterial oozingWorks best applied dry
Topical epinephrineVasoconstrictionAfter laryngeal saccule excisionUse only where the surgeon directs

Sources: Zeltzman, Veterinary Practice News2; Jamali et al.3; El-Sayed et al.1; Driessen et al.4

Gelatin sponges are reabsorbed in about four to six weeks.2 Gelatin products can absorb fluid and expand up to 45 times their dry weight, and their neutral pH means they can be combined with topical thrombin.3 Oxidized regenerated cellulose is acidic, which gives it a bactericidal effect.3 Bone wax is a nonabsorbable mixture of beeswax and paraffin; because it hinders bone healing, its use is not recommended in most cases.3

Lasers

Surgical lasers, such as the CO2 laser, cut and coagulate with focused light. The CO2 laser emits far-infrared light at 10.6 micrometers.10 In brachycephalic airway surgery, CO2 laser, monopolar electrocautery and bipolar vessel sealing devices have all been used for soft palate resection with similar outcomes.12 High-power Class IV lasers are hazardous to view directly or by scattered reflection, and they are a fire and skin hazard.10 The eye is almost always more vulnerable to laser injury than the skin.9

Procedure or technique

Setting up monopolar electrosurgery (circulating technician)

  1. Check the unit, cords and footswitch before the patient arrives. Look for cracked insulation on the pencil cord.
  2. Choose a pad site over an even, well-perfused muscle mass close to the surgical site, where fluid is unlikely to pool.1
  3. Avoid scars, hair, bony prominences, metal implants, tattoos and pressure points.1 Clip hair if needed so the whole pad touches skin.
  4. Make sure the pad lies flat with full contact. Recheck the pad after any repositioning.
  5. Prefer a non-alcohol antiseptic when an energy device is used. If alcohol is used, let it dry fully and keep it out of the hair coat.1
  6. Connect the smoke evacuator and place its intake close to where smoke is made.
  7. Set power as the surgeon directs.
  8. After surgery, remove the pad gently and inspect the skin under it. Record any redness or burn and tell the veterinarian.

Running the field (sterile technician)

  1. Keep the pencil in its holster when not in use so it cannot be activated against the patient or drapes.
  2. Open topical agents onto the field on request, and keep a count of any radiopaque sponges used.
  3. Pass hemostats ready to use and be ready to flash them when the surgeon ties a ligature.2
  4. Collect used sponges for weighing; do not discard them before the count and weight are done.

Flowchart

Checks before activating monopolar electrosurgery Surgeon will use monopolar Pad flat, full contact, over muscle, away from bone, implants, hair? Alcohol prep dry, no pooled fluid under the patient? Smoke evacuator on, oxygen not leaking near the site? Ready: tell the surgeon, holster the pencil between uses No: reseat or move pad No: wait, dry, pad No: fix first YesYesYes NoNoNo
Figure 2. Decision tree the circulating technician runs before the first activation. Every "No" ends in a fix, and the check is repeated after the patient is moved. Original diagram, VTNE Exam Prep, based on El-Sayed et al. (reference 1).

Species differences

  • Dogs: total blood volume is estimated at 80 to 90 mL/kg.7 Vessel sealers have been studied for canine laparoscopic ovariectomy.4
  • Cats: blood volume is estimated at 40 to 60 mL/kg, lower than in dogs.7 A 4 kg cat has roughly 160 to 240 mL of blood, so a few soaked sponges are a large share. Weigh sponges in small patients, not just large ones.
  • Brachycephalic dogs: palate surgery may use laser or electrosurgery in the airway.12 Oxygen is the key fire factor, so the anesthetist and surgeon must agree on oxygen and energy use before the beam or pencil is activated.1
  • Large animals: the same circuit rules apply. A large patient on a wet surface needs particular care with pad contact and pooled fluid.
Table 2. Electrosurgery and laser hazards and how the technician prevents them
HazardCauseTechnician's prevention
Burn under the return electrodePartial detachment or poor contact raises current densityFull-contact pad over muscle; recheck after moving the patient
Insulation failureDamaged insulation acts as a second active electrodeInspect cords and reusable instruments before use
Lateral thermal spreadHigh power, long activationUse the setting the surgeon orders; offer bipolar near nerves
Surgical fireOxidizer, ignition source and fuel togetherLet alcohol dry; control oxygen near the site; holster the pencil
Smoke plumeThermal destruction of tissueLocal smoke evacuation at the source
Laser eye injuryDirect or reflected Class IV beamEyewear made for that laser's wavelength for everyone in the room

Sources: El-Sayed et al.1; OSHA810

Calculation

The gravimetric method weighs used sponges and subtracts their dry weight. Blood density is close to water, so 1 g equals about 1 mL.5 Add the blood in the suction canister. The canister also holds rinse fluid,6 so subtract the lavage volume you recorded.

Worked example. A 20 kg dog has a splenectomy. Ten used lap sponges weigh 410 g in total. Each weighed 15 g dry. The suction canister holds 450 mL, and 300 mL of sterile saline was used for lavage.

Step 1, sponges: dry weight = 10 sponges × 15 g/sponge = 150 g. Blood in sponges = 410 g − 150 g = 260 g × (1 mL / 1 g) = 260 mL.

Step 2, suction: 450 mL − 300 mL lavage = 150 mL blood.

Step 3, total: 260 mL + 150 mL = 410 mL.

Step 4, share of blood volume: 20 kg × 90 mL/kg = 1,800 mL. 410 mL / 1,800 mL × 100 = about 23%.

Report the total and the percentage to the veterinarian. Weighing gives a closer estimate than looking at the sponges.

  • Surgical smoke contains carbon monoxide, polyaromatic hydrocarbons and trace toxic gases. It can irritate the upper airway and has shown mutagenic potential in the laboratory. Consensus groups recommend local smoke evacuation. OSHA has no specific standard for plume, so practice policy matters.8
  • Smoke is about 95% water vapor; the other 5% holds particles, chemicals, cells and microorganisms. Particles smaller than 2 micrometers reach the alveoli.1
  • Laser eyewear must be designed for the specific laser system in use.10 Glasses for one wavelength may not protect against another. Protect the patient's eyes as the surgeon directs.
  • Fire: oxygen, an ignition source (pencil or laser) and fuel (drapes, gauze, hair, alcohol) form the fire triad.1 Know where the fire extinguisher is and who turns the oxygen off.
  • Pacemakers: electrosurgery can interfere with implanted electronic devices and the ECG.1 Tell the surgeon if the record shows a pacemaker.
  • Client education: at discharge, describe any pad-site redness and ask the owner to report blistering or a sore patch on that area.

Common mistakes and exam traps

  • Bipolar does not need a ground plate. An answer that pairs bipolar with pad placement is wrong.
  • Rubbing a bleeding site with gauze is wrong; blot.2
  • Pad over a bony prominence or metal implant is wrong; choose a flat, well-perfused muscle mass.1
  • Gelatin sponge and cellulose control oozing. A named artery still needs ligation or sealing.
  • Do not forget to subtract dry sponge weight and lavage fluid. Forgetting either overestimates blood loss.

Quick review

  1. Blot, never rub, a bleeding field.
  2. Too much pressure stops clot formation.
  3. Flash the hemostat as the ligature is tightened.
  4. Transfixation ligatures resist slipping on large pedicles.
  5. Monopolar current returns through the ground plate; bipolar current stays between the tips.
  6. Pad: flat, full contact, over muscle, near the site, away from bone, scars, implants and hair.
  7. Vessel sealers such as LigaSure seal vessels up to 7 mm.
  8. Bone wax for cut bone; it impairs healing.
  9. Gelatin sponge swells and is reabsorbed in 4 to 6 weeks; cellulose is acidic and works best dry.
  10. Blood loss: wet minus dry sponge weight (1 g = 1 mL) plus suction minus lavage.
  11. Fire triad: oxygen, ignition, fuel; let alcohol prep dry.
  12. Use local smoke evacuation and wavelength-specific laser eyewear.

Test yourself

5 questions written for this guide. Pick an answer, then check it.

Question 1A technician is placing the return electrode for monopolar electrosurgery on a dog having a ventral abdominal mass removed. Which site is the best choice?

Show answer

Answer: B. Over a flat, clipped, well-perfused muscle mass close to the surgical site

B is right: the pad belongs over an even, well-vascularized muscle mass near the field with full contact. A: bony prominences concentrate current and pressure. C: scars are on the list of sites to avoid. D: metal implants are also to be avoided.

Source Complications of electrosurgery: mechanisms and prevention strategies, Facts Views Vis Obgyn 2024

Question 2Ten used laparotomy sponges weigh 380 g in total. Each weighed 18 g dry. The suction canister holds 260 mL, of which 200 mL was saline lavage. What is the estimated blood loss?

Show answer

Answer: B. 260 mL

B is right: sponges 380 g minus 180 g dry weight = 200 g, which is 200 mL; suction 260 mL minus 200 mL lavage = 60 mL; total 260 mL. A counts the sponges only and leaves out the canister. C adds the sponge blood to the whole 200 mL lavage volume instead of the 60 mL of blood. D ignores both dry weight and lavage.

Source Blood density is nearly equal to water density: a validation study of the gravimetric method, J Vet Med 2015

Question 3During a mandibulectomy, cut bone surfaces keep oozing. Which topical product does the surgeon most likely request to plug the bleeding bone?

Show answer

Answer: C. Bone wax

C is right: bone wax mechanically plugs channels in cut bone, though it is poorly absorbed and can impair healing. A and D are for oozing soft tissue surfaces. B causes vasoconstriction and is used, for example, after laryngeal saccule excision.

Source All bleeders stop...eventually, Veterinary Practice News

Question 4A surgeon needs to coagulate small vessels next to the spinal cord and wants current confined to the tissue between the instrument tips. What should the technician set up?

Show answer

Answer: A. Bipolar forceps, which need no return electrode

A is right: bipolar current passes only between the two tips, with lower voltage and almost no risk of alternate site burns. B and C still send current through the patient to a pad, and higher power increases lateral thermal spread. D is not used on the spine and does not coagulate.

Source Complications of electrosurgery: mechanisms and prevention strategies, Facts Views Vis Obgyn 2024

Question 5Which action best protects the surgical team during CO2 laser surgery?

Show answer

Answer: D. Wearing eyewear made for the laser's wavelength and using local smoke evacuation

D is right: laser eyewear must be designed for the specific laser system, and local smoke evacuation is recommended for plume. A spreads plume rather than capturing it. B may not block the 10.6 micrometer CO2 wavelength. C adds fuel to the fire triad.

Source OSHA Technical Manual, Section III, Chapter 6: Laser Hazards, OSHA

Practise this topic

18 questions from the practice bank are linked to this guide. Answer them in practice mode with instant explanations.

Figure credits

  • Figure 1 Original diagram, VTNE Exam Prep
  • Figure 2 Original diagram, VTNE Exam Prep

References

  1. El-Sayed SM, Saridogan E, El-Sayed MM. Complications of electrosurgery: mechanisms and prevention strategies. Facts Views Vis Obgyn. 2024;16(4). PMC11819786 pmc.ncbi.nlm.nih.gov/articles/PMC11819786/
  2. Zeltzman P. All bleeders stop...eventually. Veterinary Practice News. 2012. veterinarypracticenews.com (accessed 2026-10-31) veterinarypracticenews.com/?p=10498
  3. Jamali B, Nouri S, Amidi S. Local and systemic hemostatic agents: a comprehensive review. Cureus. 2024;16(10):e72312. PMC11585330 pmc.ncbi.nlm.nih.gov/articles/PMC11585330/
  4. Driessen F, Marrero JD, Grinwis GCM, van Nimwegen SA. Comparison of two advanced bipolar tissue sealer/dividers for laparoscopic ovariectomy in dogs: articulating Enseal G2 versus Ligasure Maryland device. Acta Vet Scand. 2023. PMC10687780 pmc.ncbi.nlm.nih.gov/articles/PMC10687780/
  5. Vitello DJ, Ripper RM, Fettiplace MR, Weinberg GL, Vitello JM. Blood density is nearly equal to water density: a validation study of the gravimetric method of measuring intraoperative blood loss. J Vet Med. 2015;2015:152730. PMC4590883 pmc.ncbi.nlm.nih.gov/articles/PMC4590883/
  6. Gerdessen L, Meybohm P, Choorapoikayil S, et al. Comparison of common perioperative blood loss estimation techniques: a systematic review and meta-analysis. J Clin Monit Comput. 2021. PMC7943515 pmc.ncbi.nlm.nih.gov/articles/PMC7943515/
  7. Merck Veterinary Manual. Blood Transfusions in Dogs and Cats. merckvetmanual.com (accessed 2026-10-31) merckvetmanual.com/circulatory-system/blood-groups-and-blood-transfusions/blood-transfusions-in-animals
  8. Occupational Safety and Health Administration. Laser/Electrosurgery Plume: Overview. osha.gov (accessed 2026-10-31) osha.gov/laser-electrosurgery-plume
  9. Occupational Safety and Health Administration. Laser Hazards: Overview. osha.gov (accessed 2026-10-31) osha.gov/laser-hazards
  10. Occupational Safety and Health Administration. OSHA Technical Manual, Section III, Chapter 6: Laser Hazards. osha.gov (accessed 2026-10-31) osha.gov/otm/section-3-health-hazards/chapter-6
  11. American Association of Veterinary State Boards. 2023 Veterinary Technician National Examination Content Outline. aavsb.org (accessed 2026-10-31) aavsb.org/wp-content/uploads/2025/04/AAVSB-2023-Veterinary-Technician-National-Examination-Content-Outline-2.pdf
  12. Today's Veterinary Practice. A Guide to Brachycephalic Obstructive Airway Syndrome Surgery. todaysveterinarypractice.com (accessed 2026-10-31) todaysveterinarypractice.com/soft-tissue-surgery/guide-to-brachycephalic-obstructive-airway-syndrome-surgery/

Spotted an error? Email support@vtneexam.com and we will check it against the sources.